Electron retardation

Term · Chemistry · MLC-T-CHM-003736

The deceleration of emitted electrons by a retarding potential, applied ahead of the energy analyser or inside it, so that their kinetic energy distribution is measured at a lowered electron energy. The analyser can then be run at a smaller pass energy, which narrows the absolute energy width, at the cost of reduced transmission. The scheme is used in Auger electron and X-ray photoelectron spectroscopy, and fixed retard ratio is a standard mode on hemispherical analysers.

Table 1. Record
IdentifierMLC-T-CHM-003736
FieldChemistry
SubjectAnalytical Chemistry
ReferencesPAC, 2020, 92, 1781. 'Glossary of methods and terms used in surface chemical analysis (IUPAC Recommendations 2020)' on page 1825 (https://doi.org/10.1515/pac-2019-0404)
See alsoAuger spectroscopy; Electron energy analyser