Brubaker lens
Term · Chemistry · MLC-T-CHM-001518
A short set of four rod electrodes mounted in line with and just ahead of the main rods of a quadrupole mass filter, driven with part of the radiofrequency voltage but without the DC voltage that the main rods carry. It passes ions above a threshold mass-to-charge ratio, and the delayed onset of the DC component limits defocusing by fringe fields, so more ions are transmitted for the same mass resolution.
| Identifier | MLC-T-CHM-001518 |
|---|---|
| Field | Chemistry |
| Subject | Analytical Chemistry |
| Synonyms | Brubaker pre-filter |
| References | PAC, 2013, 85, 1515. 'Definitions of terms relating to mass spectrometry (IUPAC Recommendations 2013)' on page 1526 (https://doi.org/10.1351/PAC-REC-06-04-06) |
| See also | Fringe field; Transmission efficiency; Transmission quadrupole mass spectrometer |